Nutriline Twinflo 2-lumen, 2Fr, 30cm met Microflash 20G en stylet

Artikelnummer: 1252.230

Nutriline Twinflo is a 2Fr polyurethane double lumen central venous catheter for premature babies and neonates inserted in a peripheral vein (basilic, cephalic, saphena...).
It is particularly recommended for medium and long-term IV therapies (up to 4 weeks) and for the simultaneous administration of incompatible drugs.

Nutriline Twinflo is inserted through a 20G peelable plastic cannula (Microflash), the needle of which features a lateral eye to quickly visualize blood reflux after successful cannulation of the vein. The peelable cannula is totally removed after catheter placement.
For the stabilization of the catheter, Nutriline Twinflo is supplied with Grip-Lok device with hypoallergenic adhesive specially designed for Neonates.

A measuring tape is supplied for the user to measure the length of catheter to insert from the puncture site to the lower third of the superior vena cava.

Product benefits:
- Double-lumen: allows simultaneous administration of incompatible drugs or solutions. Prevents trauma linked to repeated punctures.
- Catheter taper at the proximal end to minimize the risk of kinking at the junction with the fixation wings.
- Extension tubes with clamps and colour-coded hubs to easily differentiate each lumen.

5 codes are available:
1252.225: Nutriline Twinflo 20 cm without stylet and with Microflash introducer
1252.230: Nutriline Twinflo 30 cm with stylet and with Microflash introducer
1252.232: Nutriline Twinflo 30 cm without stylet and without introducer
1252.233: Nutriline Twinflo 30 cm with stylet and without Microflash introducer 
1252.235: Nutriline Twinflo 30 cm without stylet and with Microflash introducer  

Quantity box/case: 10/50


  Code Introducer Type Introducer Ext. Ø
mm
Introducer Length
mm
Introducer G
G
Catheter Length
cm
Catheter Ext. Ø
mm
Catheter Gauge
G
Catheter Prim. vol.
ml
Catheter Flow rate for 1 bar pressure
ml/min
Catheter Stylet Catheter Ø
Fr
1252.225 Microflash 1.1 18 20 20 0.6 24 0.2 (x2) 3.20 (x2) No 2
1252.230 Microflash 1.1 18 20 30 0.6 24 0.2 (x2) 1.45 (x2) Yes 2
1252.235 Microflash 20G 1.1 18 20 30 0.6 24 2 x 0.2 2 x 1.45 No 2
  Introducer Catheter
Code Type Ext. Ø
mm
Length
mm
G
G
Length
cm
Ext. Ø
mm
Gauge
G
Prim. vol.
ml
Flow rate for 1 bar pressure
ml/min
Stylet Ø
Fr
1252.225 Microflash 1.1 18 20 20 0.6 24 0.2 (x2) 3.20 (x2) No 2
1252.230 Microflash 1.1 18 20 30 0.6 24 0.2 (x2) 1.45 (x2) Yes 2
1252.235 Microflash 20G 1.1 18 20 30 0.6 24 2 x 0.2 2 x 1.45 No 2
  • Contains Latex: No
  • Contains DEHP: No
  • Contains biogical or animal-based product: No
  • Non-pyrogenic: Yes